Assembly Replacement
A35 3-20 GHz I/Q Modulator (E8267C Models Only)
A35 3-20 GHz I/Q Modulator (E8267C Models Only)
Tools Required
needle-nose pliers
Removal Procedure
Refer to Figure 2-46 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to "Outer Instrument Cover" on page 2-5
and "Inner Instrument Cover" on page 2-7.
3. Position the signal generator with the front panel facing you.
4. Using the T-10 driver, remove the four screws from the Baseband Generator cover.
5. Using the needle-nose pliers, disconnect the following flexible cables:
W63 from A35J5
W62from A35J6
W49 from A35J7
W50 from A35J8
W61 from A35J9
W60 from A3510
W59 from A35J11
W58 from A35J12
W57 from A35J13
W56 from A35J14
6. Lift the retention levers and disconnect the A35 3-20 GHz I/Q Modulator from the
A31 Motherboard.
7. Lift the A35 3-20 GHz I/Q Modulator out of the slot.
Replacement Procedure
Reverse the order of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal
procedure.

Table of Contents

Other manuals for Agilent Technologies E8267C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8267C and is the answer not in the manual?

Agilent Technologies E8267C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8267C
CategoryPortable Generator
LanguageEnglish

Related product manuals