4-7
Post-Repair Procedures
Post-Repair Procedures Matrix
Table 4-6 Performance Tests: Assemblies A34, A35, AT1, B1, and RF Output Connector
Performance Tests Replaced
Assembly
The following performance tests are listed
in the order that they should be performed
to minimize changes in test equipment
configurations.
A34 Internal Hard Drive
A35 3 20 IQ Modulator
AT1 115 dB Attenuator
B1 Fan
RF Output Connector
Self Tests
✓✓✓✓
Maximum Leveled Output Power
Power Level Accuracy
✓✓
Internal Pulse Modulation Level Accuracy
Internal Pulse Modulation Rise/Fall Time
DC FM Carrier Offset
Internal FM Frequency Response
External Phase Modulation Frequency Response
Internal FM Distortion
Internal Phase Modulation Distortion
External FM Deviation Accuracy
Internal Phase Modulation Deviation Accuracy
External Pulse Modulation ON/OFF Ratio
Harmonic Spurious
Sub-Harmonic Spurious
Non-Harmonic Spurious
Single-Sideband Phase Noise
EVM and IQ Offset
External Digital Modulation Level Accuracy
Relative to CW
Swept Frequency Accuracy
Optional Performance Test
Timebase Aging Rate
Functional Check (Service Software Utility)
Dual Arb

Table of Contents

Other manuals for Agilent Technologies E8267C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8267C and is the answer not in the manual?

Agilent Technologies E8267C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8267C
CategoryPortable Generator
LanguageEnglish

Related product manuals