1-37
Troubleshooting
Troubleshooting Assembly–Level Problems
3. If both the I and Q signals are good, reconnect the cable removed in Step 2. Check the signals at
the A8 Output board.
a. At J53-8, repeat the Q offset check in Step 1.
If the signal is good, replace the A8 Output board.
If the signal is bad, replace the A31 Motherboard.
b. At J53-19, repeat the I offset check in Step 2.
If the signal is good, replace the A8 Output board.
If the signal is bad, replace the A31 Motherboard.
405 Gain Adjustment Test
Replace the A8 Output.
406 Prelevel Bypass Test
Replace the A8 Output.
407 Switched Filters Test
Replace the A8 Output.
408 ALC Reference DAC Test
Replace the A8 Output.
409 ALC Modulation Test
Replace the A8 Output.
410 Burst Mod Driver Test
Replace the A8 Output.
411 ALC Detector Test
Replace the A8 Output.
412 Het Band Test
Measure the 1 GHz signal from the A7 Reference board:
1. Turn on the signal generator.
2. Set CW = 200 MHz.
3. Using a spectrum analyzer, probe P51-19.
There should be a 1 GHz signal, that is >30 dBm.
If the signal is good, replace the A9 Output.
If the signal is not good, run the A7 self-tests.

Table of Contents

Other manuals for Agilent Technologies E8267C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8267C and is the answer not in the manual?

Agilent Technologies E8267C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8267C
CategoryPortable Generator
LanguageEnglish

Related product manuals