1-40
Troubleshooting
Troubleshooting Assembly–Level Problems
605 Sweep Lock
1. Using an oscilloscope with high input impedance, measure the center pin of A9J9.
A positive and negative going signal should be present.
If the signal is present at A9J9, replace the A9 YIG Driver.
2. If the signal is not present at A9J9, measure the center pin of J1032 on the motherboard.
If the signal is present, replace the cable.
If the signal is not present, replace the A6 Frac–N.
Source Setup Oscilloscope
Setup
Start: 5 GHz 10 ms/div
Stop: 10 GHz 5V/div
Sweep: Freq 1 M
Sweep
Time:
Auto
Sweep
Repeat:
Continuo
us