1-57
Troubleshooting
Troubleshooting Assembly–Level Problems
1301 Internal Pulse Generator Clock
1. Using an oscilloscope, measure P131–91 for a 10 MHz signal at 3.5 V
pp
.
If the signal is good, replace the A11 Pulse/Analog Modulation Generator.
2. If the signal is bad, check P41–7.
If the signal is good on P41–7, replace the A31 Motherboard.
If the signal is bad on P41–7, replace the A7 Reference.
1302 Output
Replace A11 Pulse/Analog Modulation Generator.
1303 Voltage Ref DAC
Replace A11 Pulse/Analog Modulation Generator.
1304 20 GHz Pulse
Replace A11 Pulse/Analog Modulation Generator.
1305 Not Used
1306 Standard 3 GHz Pulse
Replace A11 Pulse/Analog Modulation Generator.
1307 High Performance 3 GHz Pulse
Replace A11 Pulse/Analog Modulation Generator.
1308 Numeric Synthesizer
Replace A11 Pulse/Analog Modulation Generator.
1309 Function Generator Channel 1
Replace A11 Pulse/Analog Modulation Generator.
1310 Function Generator Channel 2
Replace A11 Pulse/Analog Modulation Generator.
1311 Frequency Modulator
Replace A11 Pulse/Analog Modulation Generator.
1312 Low Frequency Out
Replace A11 Pulse/Analog Modulation Generator.

Table of Contents

Other manuals for Agilent Technologies E8267C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8267C and is the answer not in the manual?

Agilent Technologies E8267C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8267C
CategoryPortable Generator
LanguageEnglish

Related product manuals