4-42 Chapter4
Troubleshooting
201 to 799: Device-Specific Errors
223 Trigger Offset unavailable in swept spans
Trigger Offset is only available in Zero Span. Refer to
“Trig” on page 6-100 for a description of this function.
215 TG start freq is less than 1/2 res bw
TG uncalibrated at start frequencies below 1/2 the
current resolution bandwidth.
214 TG start freq is less than 9kHz
TG uncalibrated below 9 kHz.
204 TG Frequency Limit
The tracking generator has reached the limit of its
allowable frequency range.
736 Too many data values at <freq_or_time_value>
This message may appear when data is sent to a
corrections or limit table using the
DATA or MERGE
commands. These tables limit the number of
amplitudes associated with a frequency or time to 2 or
less. This message will appear if an attempt is made to
attach 3 or more values to a frequency or time.
206 Unable to initialize flatness data
A failure occurred in setting the flatness data in the
internal EEROM. Contact your local Agilent
Technologies sales and service office.
762 Unable to load file
A failure occurred while loading a file; the file was not
loaded.
759 Unable to load state saved from firmware
Rev A.03.00
A saved state file from a newer firmware revision was
attempted to be loaded into an older instrument.

Table of Contents

Other manuals for Agilent Technologies ESA-E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ESA-E Series and is the answer not in the manual?

Agilent Technologies ESA-E Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelESA-E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals