EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>ESA-E Series

Agilent Technologies ESA-E Series User Manual

Agilent Technologies ESA-E Series
321 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #74 background imageLoading...
Page #74 background image
3-14 Chapter3
Functional Testing
Displayed Average Noise Level
Procedure (3.01 GHz to 6.0 GHz) E4404B, E4405B,
E4407B, E4408B
1. Press System, Power On/Preset, Preset (Factory), Preset on the
analyzer. Wait for the preset routine to finish. Set the analyzer by
pressing the following keys:
FREQUENCY, Start Freq, 3.01 GHz
Stop Freq, 6.0 GHz
AMPLITUDE, –70 dBm
Attenuation (Man), 0 dB
BW/Avg, 1 MHz
Video BW Auto Man, 10 kHz
2. Press the following keys on the analyzer:
Single
BW/Avg, Average (On), 3, Enter
Single
Wait until AVG 3 is displayed to the left of the graticule (the
analyzer will take 3 sweeps and then stop).
3. Press Peak Search. Record the frequency in the Measured Frequency
column (g) in Table 3-1 for 3.01 GHz to 6.0 GHz.
4. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
BW/Avg, Average (Off)
BW/Avg, Resolution BW (Auto)
Video BW (Auto)
SPAN, 50 kHz
FREQUENCY
5. Press Center Freq. Set the center frequency of the analyzer to the
frequency recorded in the Measured Frequency column (g) in
Table 3-1 for 3.01 GHz to 6.0 GHz.
6. Press the following keys on the analyzer:
BW/Avg, 1 kHz
Video BW Auto Man, 30 Hz
Single
Wait for the sweep to finish.

Table of Contents

Other manuals for Agilent Technologies ESA-E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ESA-E Series and is the answer not in the manual?

Agilent Technologies ESA-E Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelESA-E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals