3-20 Chapter3
Functional Testing
Displayed Average Noise Level
7. Press the following keys on the analyzer:
Display, Display Line (On)
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. Record the display line amplitude setting as TR Entry 14 in
Table 3-1. The Average Noise Level should be less than the
Maximum.
Procedure (22.01 GHz to 26.5 GHz) E4407B and
E4408B
1. Press System, Power On/Preset, Preset (Factory), Preset on the
analyzer. Wait for the preset routine to finish.
Set the analyzer by pressing the following keys:
FREQUENCY, Start Freq, 22.01 GHz
Stop Freq, 26.5 GHz
AMPLITUDE, –70 dBm
Attenuation (Man), 0 dB
BW/Avg, 1 MHz
Video BW Auto Man, 10 kHz
2. Press the following keys on the analyzer:
Single
BW/Avg, Average (On), 3, Enter
Single
Wait until AVG 3 is displayed to the left of the graticule (the analyzer
will take 3 sweeps and then stop).
3. Press Peak Search. Record the frequency in the Measured Frequency
column (l) in Table 3-1 for 22.01 GHz to 26.5 GHz.

Table of Contents

Other manuals for Agilent Technologies ESA-E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ESA-E Series and is the answer not in the manual?

Agilent Technologies ESA-E Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelESA-E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals