6-16 Chapter6
Front-Panel Key Reference
BW/Avg
EMI Res BW Accesses the EMI Res BW menu keys and allows you to choose between
bandwidths of
120 kHz, 9 kHz and 200 Hz. The 200 Hz bandwidth is only
available if Option 1DR (narrow resolution bandwidth) is installed and
the span is less than 5 MHz. (Option 1DR is available with Agilent
ESA-E Series only (E4401B, E4402B, E4404B, E4405B and E4407B.)
This function is set to
None when the resolution bandwidth is set to any
other value using the
Resolution BW key.