104 1200 Series DAD and MWD User Manual
8Maintenance
Note:
Depending on the system setup, the inlet capillary
might be routed directly from the module above or
below to the cell and not to the capillary holder.
3 Disconnect the cell inlet and the cell outlet capillary and
the waste capillary from the capillary holder.
4 Loosen the thumb screw and remove the flow cell outlet
capillary (bottom) with the union.
5 Remove the flow cell while pressing the flow cell holder.