1200 Series DAD and MWD User Manual 123
Maintenance 8
Nano Flow Cell - Replacing or Cleaning
When required If parts are contaminated or leaky.
Tools required Screwdriver POZI 1 PT3
Two 1/4 inch wrenches for capillary connections
Parts required For parts identification refer to
“Nano Flow Cells” on page 158 (80 nl and
500 nl).
Preparations Turn the lamp(s) off.
Remove the front cover.
Remove the flow cell, see
Exchanging a Flow Cell” on page 103.
NOTE
For details refer to the technical that comes with the nano-flow cell kit.
NOTE
The quartz block can be cleaned with alcohol. DO NOT touch the inlet and outlet windows
at the quartz block.