124 1200 Series DAD and MWD User Manual
8Maintenance
1 Disconnect the capillaries from the capillary holder and
remove the flow cell.
2 Unscrew the cell body from the holder.
3 Unscrew the capillaries from the flow cell. DO NOT use the
adapter at this time!
4 Using for example a toothpick, press on the plastic part
and slide the quartz body out of the cell housing.