1200 Series DAD and MWD User Manual 107
Maintenance 8
1 Use a 4 mm hex key to unscrew the window assembly and
remove the gasket from the cell body.
Note:
If you want to replace the gasket only, continue with
step 7.
Do not mix the gasket #6 and # 7. They are different
for standard and semi-mirco flow cell.
2 Use a tooth pick to remove the quartz window from the
window assembly.
Note:
If the washers fall out of the window assembly, they
must be inserted in the correct order with the
TeflonĀ® ring to prevent any leaks from the flow cell
window.
6
1
1
7
Do not mix the gasket
#6 and # 7 (different
hole diameter)