110 1200 Series DAD and MWD User Manual
8Maintenance
Next steps:
10 Reconnect the capillaries, see Exchanging a Flow Cell” on page 103.
11 Perform a leak test.
12 Insert the flow cell.
13 Replace the front cover
14 Perform a
“Wavelength Verification and Recalibration” on page 138 or a “Holmium Oxide Test on page 139 to check the
correct positioning of the flow cell.