1200 Series DAD and MWD User Manual 137
Maintenance 8
Suggested Actions
✔ Ensure the flow cell is filled with water, and free from air bubbles.
✔ Run the cell test (see “Cell Test (ChemStation only)” on page 141). If the test
fails, exchange the flow cell windows.
✔ Clean optical components with alcohol and lint-free cloth.
✔ Exchange the lamp.
Figure 42 Intensity Test Results (report)