1200 Series DAD and MWD User Manual 137
Maintenance 8
Suggested Actions
Ensure the flow cell is filled with water, and free from air bubbles.
Run the cell test (see “Cell Test (ChemStation only)” on page 141). If the test
fails, exchange the flow cell windows.
Clean optical components with alcohol and lint-free cloth.
Exchange the lamp.
Figure 42 Intensity Test Results (report)