86 1200 Series DAD and MWD User Manual
5 How to optimize the Detector
The ratio range was set at 2 2.4 (2.2 ±10%). Only when the ratio between 249
and 224 nm was within this range, is the signal plotted. Of all four peaks, only
the third fulfilled the criterion (Figure 36). The others were not plotted.