Agilent InfinityLab LC Series Diode Array Detectors User Manual 216
13 LAN Configuration
Manual Configuration
4 To change a parameter follows the style:
parameter value, for example:
ip 134.40.28.56
Then press [Enter], where parameter refers to the configuration parameter
you are defining, and value refers to the definitions you are assigning to that
parameter. Each parameter entry is followed by a carriage return.
5 Use the “/” and press Enter to list the current settings.
Table 36 Telnet Commands
Value Description
? displays syntax and descriptions of commands
/ displays current LAN settings
ip <x.x.x.x> sets new ip address
sm <x.x.x.x> sets new subnet mask
gw <x.x.x.x> sets new default gateway
exit exits shell and saves all changes
Figure 67 Telnet - Current settings in "Using Stored" mode
information about the LAN interface
MAC address, initialization mode
Initialization mode is Using Stored
active TCP/IP settings
TCP/IP status - here ready
connected to PC with controller software (e.g. Agilent
ChemStation), here not connected

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7117B and is the answer not in the manual?

Agilent Technologies G7117B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7117B
CategorySecurity Sensors
LanguageEnglish

Related product manuals