Agilent InfinityLab LC Series Diode Array Detectors User Manual 217
13 LAN Configuration
Manual Configuration
6 Change the IP address (in this example 192.168.254.12) and type “/” to list
current settings.
7 When you have finished typing the configuration parameters, type
exit and press Enter to exit with storing parameters.
Figure 69 Closing the Telnet Session
Figure 68 Telnet - Change IP settings
change of IP setting to
Initialization mode is Using Stored
active TCP/IP settings
stored TCP/IP settings in non-volatile memory
connected to PC with controller software (e.g. Agilent
ChemStation), here not connected
NOTE
If the Initialization Mode Switch is changed now to “Using Stored” mode, the
instrument will take the stored settings when the module is re-booted. In the
example above it would be 192.168.254.12.