51
Order online at www.agilent.com/chem/store
Inlets
Ferrule Selection Recommendations
Ferrule/ Seal Type Upper Temp.
Limit
Usages Advantages Limitations
Graphite (100%) 450°C
General purpose for
capillary columns.
Suitable for FID and NPD.
Recommended for high
temperature and cool
on-column applications.
Easy to use stable seal.
Higher temperature limit.
Can be removed easily.
Not for MS or oxygen
sensitive detectors.
Soft, easily deformed
or destroyed.
Possible system contamination.
Vespel/Graphite (85%/15%) 350°C
General purpose for
capillary columns.
Recommended for MS and
oxygen sensitive detectors.
Most reliable leak-free
connection.
Mechanically robust.
Long lifetime.
Not reusable.
Flows at elevated temperature.
Must retighten frequently.
Vespel (100%) 280°C
Isothermal operation.
Can be reused or
removed easily.
Excellent sealing material
when making metal or glass
connections.
Mechanically robust.
Long lifetime.
Can be reused or
removed easily.
Leaks after temperature cycle.
Flows at elevated temperature.
Must retighten frequently.
SilTite (100% metal) N/A
Use with Capillary Flow
Technology nut kits
Provide leak-tight seals with
Ultimate Union, Deans Switch
and Effluent Splitter
Not reusable.
Tips & Tools
When using V
espel/Graphite ferrules, tighten the column nut 1/4 turn after
the first temperature program runs. Even preconditioned ferrules can
exhibit some shrinkage after a temperature programmed run.