EasyManua.ls Logo

Agilent Technologies LC Series

Agilent Technologies LC Series
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
108 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
6
Error Information
Module Error Messages
Pressure Below Lower Limit
Error ID: 2015, 2501
The system pressure has fallen below the lower pressure limit.
Probable cause Suggested actions
1
Lower pressure limit set too high.
Ensure the lower pressure limit is set to a value
suitable for the analysis.
2
Air bubbles in the mobile phase.
Make sure that the degasser is in flow path
and works correctly. Purge the module.
Ensure solvent inlet filters are not blocked.
3
Leak.
Inspect the pump head, capillaries and
fittings for signs of a leak.
Purge the module. Run a pressure test to
determine whether the seals or other
module components are defective.
4
Defective damper.
Please contact your Agilent service
representative.
5
Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Related product manuals