Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 55
Using the Pump
3
Best Practices
Prepare the pump
Purge
Use the Purge function to:
fill the pump,
exchange a solvent,
remove air bubbles in tubes and pump heads.
Seal wash
Seal Wash guarantees a maximum seal life time. Use Seal Wash:
When using buffers with elevated salt concentrations
When using volatile solvents with non-volatile additives
CAUTION
Contaminated seal wash solvent
Do not recycle seal wash solvent to avoid contamination.
Weekly exchange seal wash solvent.