EasyManua.ls Logo

Agilent Technologies LC Series

Agilent Technologies LC Series
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
110 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
6
Error Information
Module Error Messages
AIV Fuse
Error ID: 2044
The active-inlet valve in the module has drawn excessive current causing the
inlet-valve electronic fuse to open.
Probable cause Suggested actions
1
Defective active inlet valve.
Restart the module. If the error message
appears again, exchange the active inlet valve.
2
Defective connection cable (front panel to
main board).
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Related product manuals