144 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
8
Maintenance
Exchange the Outlet Valve
Exchange the Outlet Valve
For bio-inert modules use bio-inert parts only!
When If internally leaking
Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
8710-1924 Open-end wrench 14 mm
5067-5688 Torque wrench 1 – 25 Nm with 14 mm wrench
Parts required p/n Description
G1312-60067 Outlet valve
G5611-60067 Outlet Valve
Preparations Switch off the pump at the main power switch.
Open the doors of the module.