Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 143
Maintenance
8
Exchange the Passive Inlet Valve
3
Insert the new valve into the pump head and tighten the
valve using a torque wrench (approx. 10 Nm).
4
Reconnect the solvent inlet tube to the passive inlet
valve.
5
Close the doors.