EasyManuals Logo

Agilent Technologies LC Series User Manual

Agilent Technologies LC Series
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #118 background imageLoading...
Page #118 background image
118 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
6
Error Information
Module Error Messages
Degasser: signal fail
Error ID: 2243, 8016
The pump board gets no or wrong pressure signals from the built-in degasser:
No valid pressure signal is shown during startup of the degasser
Measured pressure is higher than 32000 hPa
Degasser: Vacuum cannot be maintained
Error ID: 8014
Pressure in degasser vacuum chamber exceeds 180 mbar limit during normal
operation.
Probable cause Suggested actions
1
Degasser sensor defect
Please contact your Agilent service
representative.
2
Degasser sensor not connected to main
board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Liquid in degasser tubing.
Please contact your Agilent service
representative.
2
Leak in degasser tubing or chamber.
Please contact your Agilent service
representative.
3
Degasser vacuum pump defective.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies LC Series and is the answer not in the manual?

Agilent Technologies LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelLC Series
CategoryWater Pump
LanguageEnglish

Related product manuals