Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 131
Test Functions and Calibration
7
Leak Rate Test
Evaluating the Results
Results of the leak rate test are the leak rates measured for pistons 1 and 2 as
described for the test principle. If any of the leak rates exceeds 3 L/min, the
test will fail.
'DPSHU
3XPSFKDPEHU
3XPSFKDPEHU
2XWOHWYDOYH
3XUJHYDOYH
,QOHWYDOYH
3LVWRQ
3LVWRQ
6HDOV
WRFROXPQ
WRZDVWH