Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 31
Introduction
1
Leak and Waste Handling
Leak Sensor
Waste Concept
1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45
with 4 ports (5043-1221) for optimal and safe waste disposal. If you decide
to use your own waste solution, make sure that the tubes don't immerse in
the liquid.
CAUTION
Solvent incompatibility
The solvent DMF (dimethyl formamide) leads to corrosion of the leak sensor. The
material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
Do not use DMF.