Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 157
Maintenance
8
Maintenance of a Pump Head Without Seal Wash Option
9
Place the pump housing onto the piston housing.
10
Insert the pistons and carefully press them into the
seals.
11
Tighten the lock screw.
12
Install the pump head, see “Reinstall the Pump Head
Assembly” on page 162.