EasyManua.ls Logo

Agilent Technologies U2000 Series

Agilent Technologies U2000 Series
101 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Specifications and Characteristics
52 U2000 Series Operating and Service Guide
Examples
U2000/1/2/4A Power Sensors
The switching point for U2000/1/2/4A sensor is –7 dBm. The hysteresis
causes the low power path to remain selected until approximately
–6.5 dBm has been reached. As the power level increases above –6.5 dBm,
the high power path will be selected. The high power path remains
selected until approximately –7.5 dBm has been reached. As the power
level decreases below –7.5 dBm, the low power path will be selected.
U2000/1/2H Power Sensors
The switching point for U2000/1/2H sensor is +3 dBm. The hysteresis
causes the low power path to remain selected until approximately
+3.5 dBm has been reached. As the power level increases above +3.5 dBm,
the high power path will be selected. The high power path remains
selected until approximately +2.5 dBm has been reached. As the power
level decreases below +2.5 dBm, the low power path will be selected.
U2000/1B Power Sensors
The switching point for U2000/1B sensor is +23 dBm. The hysteresis
causes the low power path to remain selected until approximately
+23.5 dBm has been reached. As the power level increases above +23.5
dBm, the high power path will be selected. The high power path remains
selected until approximately +22.5 dBm has been reached. As the power
level decreases below +22.5 dBm, the low power path will be selected.

Table of Contents

Other manuals for Agilent Technologies U2000 Series

Related product manuals