BYK-Gardner GmbH System Configuration | 6
295 024 855 E 2405 spectro2guide Family 61
After calibration on the external white calibration standard following message appears: “Per-
form diagnosis: Required”.
Verify performance on external test standard color (green) and external test standard gloss (high
gloss). Take care for correct alignment of instrument on standard. The instrument will guide you
through the diagnosis procedure.
NOTICE
• The auto diagnosis and calibration apply for the use with a docking station. In case of spec-
tro2go the external test and calibration standards are to be used.
• If the instrument has not been put in the docking station and no manual external auto diag-
nosis has been performed for 4 weeks, the auto diagnosis reminder will be displayed auto-
matically.