H-1.601. 00137-B3.0
Page 27 of 81
Adapter disconnencted!!
CPR
VentilateCPREventsAlarm
tidal volume
I/E
Freq
PEEP
Fio2
FIGURE 15 Alarm
PARTS PARTS
1 Upper airway pressure limit 6 Suffocation duration
2 Lower airway pressure limit 7 Upper tidal volume limit
3 Upper ventilation rate (per min.) 8 Lower tidal volume limit
limit
4 Lower ventilation rate (per min.) 9 Upper EtCO2 limit
limit
5 Upper respiration frequency limit 10 Lower EtCO2 limit
3.6
Ventilation mode Parameter linkage
relationship:
No.
Target
Parameter
Linkage
parameters
Interdependence
Note
1
Vt
Finsp
Vt= (Finsp*1000/60)*(Tinsp-Tinsp*Tplat)
Finsp
Inspiratory flow
TinspInspiration
time
TplatPlateau
time
Tplat=Platform
time ratio*T
i
2
Finsp
Vt
Vt= (Finsp*1000/60)*(Tinsp-Tinsp*Tplat)
/