Table 13: Resistance Measurement
Range Accuracy (% of reading ± floor)
Ohms low 0.00Ω - 400.0Ω 0.025% ± 0.05Ω
Ohms high 401.0Ω - 4000.0Ω 0.025% ± 0.5Ω
Table 14: Resistance Source
Range (IE)Excitation Current Accuracy (% of reading ± floor)
Ohms low 5.0Ω - 400.0Ω 0.1mA - 0.5mA 0.025% ± 0.03/IEΩ
5.0Ω - 400.0Ω 0.5mA - 3mA 0.025% ± 0.05Ω
Ohms high 400Ω - 1500Ω 0.05mA - 0.8mA 0.025% ± 0.5Ω
1500Ω - 4000Ω 0.05mA - 0.4mA 0.025% ± 0.5Ω
Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <= 5ms.
Table 15: Thermocouple Measurement/Source
Range Accuracy (% of reading ± floor)
Read (mV) -10.000mV - 75.000mV 0.02% ± 10µV
Source (mV) -10.000mV - 75.000mV 0.02% ± 10µV
Maximum current output in voltage ranges is 1mA with an output impedance of <= 1Ω
Table 16: Thermocouple Read and Source (errors in °C)
49
TC Type Range (°C) Accuracy
J -210.0 - 0.0 0.6
0.0 - 800.0 0.4
800.0 - 1200.0 0.5
K -200.0 - 0.0 0.8
0.0 - 1000.0 0.5
1000.0 - 1372.0 0.7
T -250.0 - 0.0 0.8
0.0 - 400.0 0.4
E -250.0 - -100.0 0.8
-100.0 - 1000.0 0.4
R 0.0 - 1767.0 1.4
S 0.0 - 1767.0 1.4
B 600.0 - 800.0 1.4
800.0 - 1000.0 1.5
1000.0 - 1820.0 1.7