Table 17: RTD Read and Source
RTD Type Range (°C) Accuracy
P10(90)385 -200.0 - 100.0 1.4
100.0 - 300.0 1.6
300.0 - 600.0  1.8
600.0 - 800.0 2.0
P50(90)385 -200.0 - 100.0 0.4
100.0 - 300.0 0.5
300.0 - 600.0  0.6
600.0 - 800.0 0.7
P100(90)385 -200.0 - 100.0 0.2
100.0 - 300.0 0.3
300.0 - 600.0  0.4
600.0 - 800.0 0.5
P200(90)385 -200.0 - 100.0 0.8
100.0 - 300.0 0.9
300.0 - 630.0  1.0
P400(90)385 -200.0 - 100.0 0.4
100.0 - 300.0 0.5
300.0 - 630.0  0.6
50
TC type Range (°C) Accuracy
C 0.0 - 1000.0 0.8
1000.0 - 2316.0 2.5
XK -200.0 - 800.0 0.4
BP 0.0 - 800.0 1.1
800.0 - 2500.0 2.5
L -200.0 - 0.0 0.45
0.0 - 900.0 0.4
U -200.0 - 0.0 0.7
0.0 - 600.0 0.45
N -200.0 - 0.0 1.0
0.0 - 1300.0 0.6
All TC errors include CJC errors
CJC error outside of 23 ± 5°C is 0.05°C/°C