EasyManua.ls Logo

Ametek JOFRA ASC300 - Page 52

Ametek JOFRA ASC300
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 17: RTD Read and Source
RTD Type Range (°C) Accuracy
P10(90)385 -200.0 - 100.0 1.4
100.0 - 300.0 1.6
300.0 - 600.0 1.8
600.0 - 800.0 2.0
P50(90)385 -200.0 - 100.0 0.4
100.0 - 300.0 0.5
300.0 - 600.0 0.6
600.0 - 800.0 0.7
P100(90)385 -200.0 - 100.0 0.2
100.0 - 300.0 0.3
300.0 - 600.0 0.4
600.0 - 800.0 0.5
P200(90)385 -200.0 - 100.0 0.8
100.0 - 300.0 0.9
300.0 - 630.0 1.0
P400(90)385 -200.0 - 100.0 0.4
100.0 - 300.0 0.5
300.0 - 630.0 0.6
50
TC type Range (°C) Accuracy
C 0.0 - 1000.0 0.8
1000.0 - 2316.0 2.5
XK -200.0 - 800.0 0.4
BP 0.0 - 800.0 1.1
800.0 - 2500.0 2.5
L -200.0 - 0.0 0.45
0.0 - 900.0 0.4
U -200.0 - 0.0 0.7
0.0 - 600.0 0.45
N -200.0 - 0.0 1.0
0.0 - 1300.0 0.6
All TC errors include CJC errors
CJC error outside of 23 ± 5°C is 0.05°C/°C