32 26-03-2004 123200 03
2. Remove the screw and clamp holding the reference
sensor in position.
3. Pull out the old reference sensor and replace it with the
new spare part 105291.
4. Insert the new reference sensor in the same position as
the old sensor.
5. Ensure that the two wires in the bend between the
ceramic tubes do not short-circuit.
6. Replace the clamp and the screw as before.
For CTC-1200 A:
If the reference- or overtemperature sensor is defect the
entire well unit must be replaced. Follow steps A through D
for removal of well unit.
Note…
Instruments should be reassembled in reverse order. Ensure
that all parts are in good working order and that all wires are
connected correctly.