EasyManua.ls Logo

AmpliVox Otowave 302+ - Ear Seal Checks

AmpliVox Otowave 302+
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USING THE OTOWAVE 302+
32 OTOWAVE 302+ INSTRUCTION FOR USE
STANDARD EAR SEAL CHECK
EXTENDED EAR SEAL CHECK
The default STANDARD option only shows if a seal could
be obtained.
The EXTENDED EAR SEAL CHECK shows a number of bars
indicating the level of the seal.
The probe should be adjusted in the ear until two or more
bars are shown for LOW and HIGH.
LOW: The pump is moved by a fixed distance in an
attempt to reduce the pressure in the ear and is held in
that position. If the measured pressure reduces by a
minimum amount, and remains low (within present
tolerances), the seal is considered OK.
HIGH: The process is repeated, at a pressure higher than
ambient. Otherwise, the process restarts. If the higher
pressure is sufficiently above the ambient pressure, and is
held, the seal is good, and the pump goes just above the
starting pressure
The method used for the extended ear seal check places
a maximum limit on the ear canal volume of ~4.5 ml.
As soon as the pressure can be established, the
tympanometry measurement will be performed. On
screen, you can follow the real-time tympanogram.
During the actual testing phase, the indicator LED b on the device and the indicator light on the probe will pulse green.
When the test sequence is completed, LED b on the device and
the indicator light on the probe will be a solid green colour. Also,
the instruction on the screen will ask you to remove the probe
from the patient’s ear.

Table of Contents

Related product manuals