TECHNICAL SPECIFICATIONS
50 OTOWAVE 302+ INSTRUCTION FOR USE
printed and/or sent to a computer, parameters used for
analysis, 128 bit Globally Unique Identifier (GUID)
Records listed in reverse chronological order (latest first),
with indication of data stored as described above
Tympanogram, Tympanogram analysis parameters, Reflex
graphs, Reflex analysis parameters, Serial Number of device,
Last and Next Due Calibration dates; space for patient &
clinician’s details to be entered.
Patient header, full left & right ear data.
226 Hz ±2% and 1000 Hz ±2%
85 dB SPL ±2dB and 79 dB SPL ±2 dB over ECV range
+200 daPa to -400 daPa ±10daPa or ±10 % (whichever is
larger) over range 0.1 ml to 6 ml
Selectable: 100, 200 or 300 daPa/sec
Admittance peak level (in ml or mƱ) pressure at peak;
Gradient in daPa (for 226Hz);
Ear Canal Volume (ECV) @ 200 daPa or -400 daPa
7.4. ACOUSTIC REFLEX TESTING
500 Hz, 1 kHz, 2 kHz & 4 kHz (±2 %)
70 dBHL to100 dBHL (±3 dB)
(2 kHz level is restricted to maximum 95 dBHL for ear canal
volumes greater than ~3.5ml)
(4 kHz level is restricted to maximum 85 dBHL for ear canal
volumes greater than ~3.5 ml & maximum 95 dBHL for all ear
canal volumes)
Number of reflex levels
presented below the selected
maximum and step size(s)
available:
100dBHL max, with 5dB or 10dB steps
95/90/85dBHL max, with 5dB steps
500Hz, 1kHz, 2kHz & 4kHz (±2%)
70dBHL to110dBHL (±3dB)
(1kHz level is restricted to minimum 75dBHL for ear canal
volumes less than ~0.2ml)
(4kHz level is restricted to maximum 100dBHL for ear canal
volumes greater than ~3.5ml)