TECHNICAL SPECIFICATIONS
51 OTOWAVE 302+ INSTRUCTION FOR USE
Number of reflex levels
presented below the selected
maximum and step size(s)
available:
110/105/100dBHL max, with 5dB or 10dB steps
95/90/85dBHL max, with 5dB steps
General
THD:
< 5 %
Reflex analysis
Reflex pass/fail at each level tested; maximum amplitude of
each reflex; nominal pressure used for the reflex test
(computer display only)
Pressure used for reflex
measurement
Pressure at tympanogram peak (if found) or at 0daPa
Reflex stimulus control
Stimulus presented at all levels, or
Stimulus ceases when a reflex is found
Reflex detection threshold
and accuracy
0.01ml to 0.5ml ±0.01ml (configurable in 0.01ml steps)
Reflex tone duration
0.6 seconds