EMC GUIDANCE & MANUFACTURER’S DECLARATION
53 OTOWAVE 302+ INSTRUCTION FOR USE
Guidance and manufacturer’s declaration – electromagnetic emissions
The Otowave 302+ Tympanometer is intended for use in the electromagnetic environment specified below. The
customer or user of the Otowave 302+ Tympanometer should assure that it is used in such an environment.
Electromagnetic environment – guidance
The Otowave 302+ Tympanometer uses RF energy only
for its internal function. Therefore, its RF emissions are
very low and are not likely to cause interference in
nearby electronic equipment.
The Otowave 302+ Tympanometer is suitable for use in
all establishments, other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Guidance and manufacturer’s declaration – electromagnetic immunity (1)
The Otowave 302+ Tympanometer is intended for use in the electromagnetic environment specified below. The
customer or user of the Otowave 302+ Tympanometer should assure that it is used in such an environment.
Electromagnetic environment – guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
±1 kV
differential
mode
±2 kV common
mode
±1 kV
differential
mode
±2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.