EasyManua.ls Logo

AND MPA-10 - Precautions before Dispensing

Default Icon
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
7-6 Precautions before dispensing
1) Before dispensing any material ensure the power is on. If power is switched on
when tip is submerged in a liquid the piston will return and the tip holder will become
wet.
2) Precautions for using the MPA-10000
To prevent liquid from touching the tip holder be sure to attach the MPA-10000 filter
to the tip holder.
When removing the tip be sure not to twist as the assembling screw in the lower
part may become loose.
3) If the tip cannot be removed via the eject button consult “12 ADJUSTING TIP
EJECTOR HEIGHT
4) Changing the number of MPB channels
The MPB can be used with 1 to 8 cannels depending on the number of cylinder
units installed.
Please refer to 13. Storage and maintenance for how to remove the cylinder unit.
Filter
MPA-10000 Lower part
Tip holder
Example with 4 channels
Cylinder unit

Table of Contents