EasyManua.ls Logo

AND MPA-10 - Mixing Mode (MIX)

Default Icon
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
When carrying out pre-rinse for multiple dispensing or stopping multiple dispensings,
use the total discharge function. (Refer to “9-8. Total discharge function”)
Minimum dispensing amount and maximum dispensing count for multiple dispensing
mode is as follow.
MODEL Minimum dispensing amount Maximum dispensing count
MPA-10 0.3μL 33 times
MPA-20 0.3μL 66 times
MPA-200 3μL 66 times
MPA-1200 15μL 80 times
MPA-10000 0.1mL 99 times
MPB-200-8 3μL 66 times
MPB-1200-8 15μL 80 times
2) Selecting the multiple dispensing mode
(The display is an example.)
[ 1 ] Press the Back key.
[ 2 ] Press the Up or Down key to select ”MD”.
[ 3 ] Press the Enter key to select multiple
dispensing mode (MD).
[ 4 ] Press the Up or Down key to set dispensing
amount for one time.
[ 5 ] Press the Enter key to confirm dispensing
amount for one time.
[ 6 ] Press the Up or Down key to set dispensing
count.
* “Dispensing amount for one time x dispensing
count” can not be set if it exceeds the volume range.

Table of Contents