Shamrock 500i & 750
21
Version 2.1 rev 13 Feb 2017
3.1.1
removing the flipper mirror locking ScrewS
The ipper mirror(s) used to switch between inputs/outputs are secured by locking screws to prevent movement during
transport.
These locking screws must be removed before turning on your Andor Shamrock spectrograph.
No power should be applied to the spectrograph at this stage.
500i Model
If your Shamrock 500 has a dual input option, there are locking screws located as shown in the following gures:
Figure 5: Location of the Shipping Screws (Shamrock 500i, Dual input Model) (1)
If your Shamrock 500 has a dual output option, there is a screw located as shown below:
Figure 6: Location of the Shipping Screw (Shamrock 500i, Dual output Model) (2)
Flipper
mirror
locking
screw
Flipper
mirror
locking
screw