pre-aligned camera and Spectrograph
When a pre-aligned camera and spectrograph package has been purchased the camera comes pre-tted with a
ange. This ange accurately locates the camera and ange assembly to the spectrograph body via precision pins and
matching bushings (see gure below). The camera and ange assembly can be easily be removed and re-attached
without the need for additional calibration, or readjustments.
Figure 9: Attaching a Pre-aligned Camera and Shamrock Spectrograph
3.2.3.1
attaching the camera
1. Insert the alignment pins into the relevant locators.
2. Secure using the four M4x12 mm screws.
3. Once connected, no further adjustment is necessary as the unit is aligned and calibrated at the factory.
Secure using M4 x 12
Screws (x4)
Insert using the
alignment pins