SECTION 5: MAINTENANCE
WARNINGS
•  THERE ARE NO USER-SERVICEABLE PARTS INSIDE THE SHAMROCK.
•  A NUMBER OF SCREWS ON THE SHAMROCK HAVE BEEN MARKED WITH RED ANTI-TAMPER PAINT. IF 
THESE ARE REMOVED, YOUR WARRANTY WILL BE VOID.
•  IF SERVICE IS REQUIRED PLEASE CONTACT YOUR ANDOR REPRESENTATIVE.
•  THE LID SHOULD ONLY BE OPENED WHEN ABSOLUTELY NECESSARY, E.G. WHEN INSTALLING A NEW 
GRATING TURRET.
•  TO REPLACE OR INSTALL OTHER ACCESSORIES PLEASE CONTACT ANDOR CUSTOMER SUPPORT 
FOR FURTHER INSTRUCTIONS.
5.1   regular checkS
•  The state of the product should be checked regularly, esp. the integrity of the External Power Supply and the 
mains cable. 
•  Do not use equipment that is damaged
5.2   annual electrical Safety checkS
•  It is advisable to check the integrity of the insulation and protective earth of the product on an annual basis, e.g. 
U.K. PAT testing.
5.3   fuSe replacement
There is a fuse (5A) located in a fuse holder to the right of the power input socket (see Section 2.2.2). If this needs to be 
replaced replace with type of same type and value.
In the U.K, Ireland and some other countries, the supplied mains cable has a BS 1363 (or Type G) plug that includes an 
integrated fuse. The characteristics of a replacement fuse are as follows:
•  Rated Current: 5 A
•  Rated Voltage: 240 VAC
•  Size:  ¼  × 1” (6.3 × 25.4 mm) cartridge
•  Type: BS 1362
5.4   cleaning (external)
•  If the external surfaces need cleaned, remove power from the unit and use a water diluted mild detergent to 
lightly dampen the cloth - do not use Isopropyl alcohol, solvents or aerosol based cleaning solutions.
•  Only use a clean, lint free cloth to clean the external painted surfaces.