Shamrock 500i & 750
55
Version 2.1 rev 13 Feb 2017
2. Click the Optimise button.
If you have Peak Checking enabled the following dialog will pop up at each of your chosen wavelength points
to conrm the correct peak is being displayed.
Two red lines will be drawn around the center range being used and the user should conrm that the peak they
are expecting is within this range.
If the user selects Yes the optimisation will continue.
If not in range the user can select Adjust Offset.
By adjusting the grating offset the user can move the peak and try to move it in range.
The user can select Cancel at any time and this will abort the current optimisation procedure.
During the acquisition the user can choose to abort and the operation will terminate at the earliest opportunity.
3. Choose to save (or not save) the values to the shamrock EEPROM.