ML248xA / ML249xA Assembly and Connections
3-14 13000-00162
September 2005
frequency. The correct calibration factor for this frequency is automatically
interpolated and applied when in V/GHz calibration factor mode. Different scaling
may be applied to sensor A or B allowing for measurement of frequency translation
devices. Available simultaneously with channel A and/or B data, the data rate is as
set on the channel. Also used for PAE measurements using a current probe.
I Output 2
Multi-purpose BNC connector is user configurable for Analog Output 2
(volts/units),or Limits Pass/fail (TTL). Supports pass/fail testing for channel 2. Also
configurable to output a real-time measurement signal from sensor input B, suitable
for levelling purposes. Also supports trigger out.
J AC main power input
85-264 VAC, 47- 440 Hz, 80 VA maximum. Connecting the AC power puts the
instrument into standby mode.