Cleaning the dentist's board.
Clean the dentist's board using a suitable product (see paragraph 1.4).
x Removable instrument holder; just slide it out of its seat to remove.
m Board handle that can be removed and sterilized in an autoclave at up to
121°C.
To remove the handle, first press the special locking buttons.
v Silicone protection that can be sterilized in an autoclave at up to 135°C.
Removable instrument hoses.
Clean and sanitise the external surface of the instrument hose using a
suitable product (see paragraph 1.4).
Instrument hoses are NOT designed for sterilization in
autoclave or by being cold soaked in solution.
Boards, version with Upper Recovery: to remove hoses, first remove
instrument holder (x).
Before removing instrument hoses, turn dental unit off.
After having turned dental unit off, drain syringe ducts by pressing the relevant air and water buttons directly on bowl until all spray
water has flowed out.
TURBINE, MICROMOTOR and SCALER instrument hoses contain water; it is therefore recommended to remove the hose by keeping
the handpiece side end on the bowl.
When refitting a hose, make sure that electric contacts are dry and that the plastic ring nut is well tightened.
Each hose must be refitted only and exclusively inside the corresponding instrument holder.