Handpiece status
An optical guide, illuminated by a multicolour LED indicator, found in the area near the control button (g), shows handpiece status as per the table given
below:
Colour
Situation
Blue light flashes, very slowly
Handpiece in standby
Light blue steady light
Handpiece activated, live images
displayed
Blue/ light blue flashing light
Handpiece in image freeze mode
Brief red flashes
Internal error: contact Technical
Service
MyRay iCapture.
This programme allows the C-U2 camera to be set up when it is connected
to a PC/WORKSTATION.
For a full description on how the MyRay iCapture programme works, refer
to the instructions, in electronic format, supplied with the C-U2 handpiece.
Disposable infection control sheaths.
The camera can be a source of cross-contamination between patients.
For this reason always use it with a disposable infection control sheath
(code 97901590) and disinfect it on the outside after use every day.
The sheath (with white paper backing) is enclosed in two protective layers:
a transparent one with blue tab at the front and a paper one at the back.
Follow the directions below to install a new disposable infection control
sheath:
Insert the camera handpiece tip between the layer with white tab and
the rear paper backing. The lens, surrounded by the LEDS, must face
down, towards the rear paper layer. Gently push the handpiece to the
end of the sheath.
Pull the blue tab removing the protective covers.
The dental camera is now protected and ready for use.
Always make certain the handpiece is correctly inserted inside the infection control sheath.
To ensure user and patient safety, always change the disposable infection control sheath before using the device on a new patient.
Disposal: the disposable infection control sheaths are to be treated as special waste materials (like surgical gloves).
Cleaning and disinfection.
Clean the handpiece with a suitable product after each use: refer to paragraph 1.5.
The camera is not designed for cold immersion sterilization.
All products must be used as directed by the manufacturer.
Maintenance and repairs.
The C-U2 dental camera does not require any particular maintenance. In the event of malfunctions, please send back the complete handpiece.
There are no parts that can be repaired on site. In the event of a malfunction, please contact an authorised dealer.
Returning parts.
Please send back any defective devices in their original packaging. Do not reuse damaged boxes.
The device must be disinfected before being shipped to prevent cross-contamination. Handpieces that have not been adequately cleaned and disinfected
will not be accepted.
The sender shall be held responsible for any equipment damage occurred during shipment regardless of whether or not the devices
are under warranty.