EasyManuals Logo

Apera Instruments PC 8500 User Manual

Apera Instruments PC 8500
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
14
6.2.2. Calibration intervals
(a) The meter is calibrated before leaving the factory and can be used right out of the box.
(b) Perform calibration once per month.
(c) For high accuracy measurements or large temperature deviations from the reference temperature
(25°C), perform calibration per week.
(d) Use conductivity standard solution to check whether there is error. Perform calibration for large errors.
(e) For new electrode or factory default setting, perform 3-point or 4-point calibrations. Choose closer
standard solution to the sample solution to perform 1- point or 2-point calibration.
6.2.3. Reference temperature
Reference temperature of factory default is 25°C. Other reference temperature can also be set for range
15°C 30°C. Select in parameter setting P2.3 and see clause 7.4.
6.2.4. Temperature coefficient
The temperature compensation coefficient of the meter setting is 2.00%/°C. However, the conductivity
temperature coefficient is different for solutions of a different variety and concentration. Please refer to chart
10 and the data collected during testing. Set per parameter P2.4. and see clause 7.4.
Notes:
When the coefficient for the temperature compensation is set to 0.00 (no compensation), the
measurement value will be based on the current temperature.
Chart 10 Temperature compensation coefficient of special solutions
Solution
Temperature compensation coefficient
NaCl salt solution
2.12%/°C
5%NaOH solution
1.72%/°C
Dilute ammonia solution
1.88%/°C
10% hydrochloric acid solution
1.32%/°C
5% sulfuric acid solution
0.96%/°C
6.2.5. Avoid contamination of standard solution
Conductivity standard solution has no buffer. Please avoid any possible contamination during usage.
Submerge the electrode in standard solution before washing the electrode and allow it dry. Please renew
conductivity standard solution frequently especially for standard solution of low concentration 84 μS/cm. The
contaminated standard solution can affect accuracy.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Apera Instruments PC 8500 and is the answer not in the manual?

Apera Instruments PC 8500 Specifications

General IconGeneral
BrandApera Instruments
ModelPC 8500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals