2. Specifications
2.1. Main specifications
Models
pH
Measuring Range
(-2.00 ~ 16.00) pH
PH8500
PC8500
Resolution
0.1/0.01 pH
Accuracy
±0.01pH ±1 digit
Temperature
compensation
(0 ~ 100) °C (manual or automatic)
mV
Measuring Range
±1,999mV
Resolution
±200 mV:0.1mV;others:1 mV
Accuracy
±0.1% F.S ±1 digit
Conductivity
Measuring Range
Conductivity0~200 mS/cm
divided into five ranges:
(0.00~19.99) μS/cm; (20.0~199.9) μS/cm;
(200~1999) μS/cm; (2.00~19.99) mS/cm;
(20.0~199.9) mS/cm
TDS: (0~100) g/L, Salinity: (0~100) ppt
EC8500
PC8500
Resolution
0.01/0.1/1μS/cm 0.01/0.1mS/cm
Accuracy
±1.0% F.S ±1 digit
Temperature
compensation
(0 ~ 100) °C (manual or automatic)
Electrode constant
0.1 / 1 / 10cm
-1
Temperature
Measuring Range
-10~110°C
PH8500
EC8500
PC8500
Resolution
0.1
Accuracy
±0.5±1 digit
2.2. Other specifications:
Data storage
500 groups
Storage content
Serial number, date, time, measuring value,
measuring unit and temperature value
Output
USB
Power
AA batteries × 3 (1.5V× 3)
IP rating
IP57 waterproof and dust proof
Dimension & Weight
Meter: (86×196×33 ) mm / 335 g
Portable case: (330×270×82 ) mm /1.3 kg
PH8500
EC8500
Portable case: (370×270×77 ) mm /1.8 kg
PC8500