6.6.4. Factory default setting
For factory default setting, please refer to parameter P2.8 (clause 7.4). All calibration data is deleted and the
meter restores to the theory value. Some functions restore to the original value (refer to appendix -1). When
calibration or measurement fails, please restore the meter to factory default setting and then perform
re-calibration or measurement. Please note once factory default setting is set, all the data deleted will not
irretrievable.
6.7. Conductivity electrode maintenance
6.7.1. Always keep the conductivity electrode clean. Before taking a measurement, rinse the electrode in
pure water and then rinse it in the sample solution. When submerging the electrode in solution, stir the
solution briefly to eliminate air bubbles and allow it to stay until a stable reading is reached. For conductivity
electrode which keeps dry, soak the electrode in pure water for 5-10 minutes. Rinse the electrode in pure
water after measurement.
6.7.2. The interaction pole of Model 2301T-S conductivity electrode is coated with platinum black to minimize
electrode polarization and expand measuring range. The platinum black coating applied by a special
processing technology, which improves the electrode performance and the robustness of the coating, If the
platinum black electrode is stained, gently clean the electrode with a soft brush in warm water containing
detergent or alcohol.
7. Parameter setting
7.1 Main menu
In the measurement mode, press key to enter in P1.0, then press or to switch to main menu:
P1.0→P2.0→P3.0. Please refer to Diagram – 14.
P1.0: pH parameter setting menu,
P2.0: Conductivity parameter setting menu,
P3.0: Basic parameter setting menu.